基于聚偏氟乙烯传感器的心冲击信号影响因素分析
作者:
作者单位:

作者简介:

通讯作者:

基金项目:

国家自然科学基金项目(61773110、61801104);辽宁省自然科学基金项目(20170540312);中央高校基本科研业务费项 目(N181604006)

伦理声明:



The Analysis of the Influence Factors on the Acquisition of Ballistocardiography Signal Based on Polyvinylidene Fluoride Sensor
Author:
Ethical statement:

Affiliation:

Funding:

  • 摘要
  • |
  • 图/表
  • |
  • 访问统计
  • |
  • 参考文献
  • |
  • 相似文献
  • |
  • 引证文献
  • |
  • 资源附件
    摘要:

    心冲击图(Ballistocardiography,BCG)记录了人体心脏做机械运动时对外界的作用力信号,不 仅能反应心脏节律,还能作为评估心脏血流动力学参数变化的有效依据。但 BCG 信号的采集过程中影 响因素较多,这些严重地影响着信号形态特征,进而导致对其进行血流动力学分析得不准确。因此, 该文采用聚偏氟乙烯压电薄膜传感器,设计并实现了能够有效抑制外界干扰、具有高信噪比的 BCG 信 号采集系统。基于该系统研究了传感器位置、床垫硬度、躺卧姿势对采集系统稳定性的影响,并分析 了在呼吸暂停情况下采集系统的检测性能。结果表明,在卧姿采集 BCG 信号时,应选择硬质床垫(如 木板床),并将传感器摆在心脏正下方的位置,使受试者保持平躺姿势。满足上述条件时,所采集的 BCG 信号波形最佳,H、J、K 波波形明显,且具有较强的节律性,有利于提高后续血流动力学分析的 精度。

    Abstract:

    Ballistocardiography (BCG) is a technique that records the force signal caused by the ejection of blood into the arteries with each heartbeat. BCG signal can be acquired non-invasively by using polyvinylidene fluoride (PVDF) sensors integrated in a mattress or chair. In addition to being able to show cardiac rhythms, BCG signals can also be used for assessing cardiac hemodynamic changes. However, body movements, sensor displacement, etc. can easily influence the morphological characteristics of BCG signal, which makes its hemodynamic analysis difficult and even impossible. In this paper, we developed a BCG signal acquisition system using PVDF sensor. The proposed system can effectively suppress external electromagnetic interference and result in high signal-to-noise ratio. Specifically, we systematically investigated the effects of sensor position, mattress hardness and user’s posture on the stability of the BCG acquisition system, and analyzed the detection performance under apnea and the difference in gender characteristics of BCG signals. The evaluation results demonstrate that when collecting BCG signals in a lying position, it is better to use a hard mattress (such as a wooden bed) and place the sensor directly below the hear. When the above conditions are met, the collected BCG signals waveform is the optimal; the H, J, and K waves are significant and have a strong rhythm, which is conducive to improving the accuracy of hemodynamic analysis.

    参考文献
    相似文献
    引证文献
引用本文

引文格式
王 璐,王仲怡,吴子悦,等.基于聚偏氟乙烯传感器的心冲击信号影响因素分析 [J].集成技术,2020,9(1):36-44

Citing format
WANG Lu, WANG Zhongyi, WU Ziyue, et al. The Analysis of the Influence Factors on the Acquisition of Ballistocardiography Signal Based on Polyvinylidene Fluoride Sensor[J]. Journal of Integration Technology,2020,9(1):36-44

复制
分享
文章指标
  • 点击次数:
  • 下载次数:
  • HTML阅读次数:
历史
  • 收稿日期:
  • 最后修改日期:
  • 录用日期:
  • 在线发布日期: 2020-01-17
  • 出版日期:
Baidu
map